2. PRESS ENTER- RESETS UPER DISPLAY TO ZERO
AND ENABLES UNIT TO COUNT AT 10 TIMES
NORMAL COUNT RATE.
3. RUN SUFFICIENT MATERIAL ACROSS SCALE FOR
ACCURATE TEST.
4. PRESS ENTER TO ACKNOWLEDGE END OF TEST.
5. PRESS ENTER TO STORE DISPLAYED AMOUNT.
6. PRESS SET-UP, 1,7,ENTER. DISPLAYS AMOUNT
STORED IN PREVIOUS STEP.
7. ENTER ACTUAL WEIGHT OF TEST LOAD.
8. PRESS ENTER
9. RECORD UPPER DISPLAY AS LEFT SPAN.
10.OBTAIN NEW CAL CONSTANT FOR FUTURE CAL
TESTS BY COMPLETING THE FOLLOWING STEPS.
11.PRESS AUTO SPAN- LIGHT FLASHES. OBSERVE
LEFT-HAND UPPER DISPLAY FOR “EL” ELECTRONIC
CAL OR ABSENCE OF “EL” FOR TEST WEIGHT OR
CHAIN CAL. IF WRONG MODE IS SELECTED, PRESS
SET-UP,1,5, ENTER FOLLOWED BY O,ENTER,AUTO
SPAN FOR TEST WEIGHTS OR CHAIN OR
1,ENTER,AUTO SPAN FOR ELECTRONIC CAL.
12. APPLY TEST WEIGHTS OR CHAIN IF THIS
METHOD IS TO BE USED.
13. PRESS AUTO SPAN, LIGHT STEADY
14. WAIT FOR ENTER LIGHT AND UPPER DISPLAY TO
FLASH.
15. RECORD UPPER DISPLAY FOR NEW CAL
CONSTANT. DO NOT PRESS ENTER.
16. PRESS SET-UP, 6,ENTER.
17. INSTALL NEW CAL CONSTANT USING THE
KEYPAD.
18. PRESS ENTER- UPPER DISPLAY WILL BLANK FOR
½ SECOND THEN DISPLAY ENTERED VALUE.
19. IF USING STATIC WEIGHTS OR CHAIN, REMOVE
19