Download manuales
Transcript
ScaleResource.com 2. PRESS ENTER- RESETS UPER DISPLAY TO ZERO AND ENABLES UNIT TO COUNT AT 10 TIMES NORMAL COUNT RATE. 3. RUN SUFFICIENT MATERIAL ACROSS SCALE FOR ACCURATE TEST. 4. PRESS ENTER TO ACKNOWLEDGE END OF TEST. 5. PRESS ENTER TO STORE DISPLAYED AMOUNT. 6. PRESS SET-UP, 1,7,ENTER. DISPLAYS AMOUNT STORED IN PREVIOUS STEP. 7. ENTER ACTUAL WEIGHT OF TEST LOAD. 8. PRESS ENTER 9. RECORD UPPER DISPLAY AS LEFT SPAN. 10.OBTAIN NEW CAL CONSTANT FOR FUTURE CAL TESTS BY COMPLETING THE FOLLOWING STEPS. 11.PRESS AUTO SPAN- LIGHT FLASHES. OBSERVE LEFT-HAND UPPER DISPLAY FOR “EL” ELECTRONIC CAL OR ABSENCE OF “EL” FOR TEST WEIGHT OR CHAIN CAL. IF WRONG MODE IS SELECTED, PRESS SET-UP,1,5, ENTER FOLLOWED BY O,ENTER,AUTO SPAN FOR TEST WEIGHTS OR CHAIN OR 1,ENTER,AUTO SPAN FOR ELECTRONIC CAL. 12. APPLY TEST WEIGHTS OR CHAIN IF THIS METHOD IS TO BE USED. 13. PRESS AUTO SPAN, LIGHT STEADY 14. WAIT FOR ENTER LIGHT AND UPPER DISPLAY TO FLASH. 15. RECORD UPPER DISPLAY FOR NEW CAL CONSTANT. DO NOT PRESS ENTER. 16. PRESS SET-UP, 6,ENTER. 17. INSTALL NEW CAL CONSTANT USING THE KEYPAD. 18. PRESS ENTER- UPPER DISPLAY WILL BLANK FOR ½ SECOND THEN DISPLAY ENTERED VALUE. 19. IF USING STATIC WEIGHTS OR CHAIN, REMOVE 19