Download GPIB Manual - wavecrest dts

Transcript
• PARAMETER:FILTER:MAXIMUM
The PARAMETER:FILTER:MAXIMUM command selects the maximum filter time in seconds.
The PARAMETER:FILTER:MAXIMUM query returns the maximum filter value.
Command syntax- :CYCLetocycle:PARAMeter:FILTer:MAXimum<-2.5 to 2.5>
Example:
Send(0,5,":CYCL:PARAM:FILT:MAX -2.5",25,EOI);
Query syntax- :CYCLetocycle:PARAMeter:FILTer:MAXimum?
Example:
Response:
Example:
Send(0,5,":CYCL:PARAM:FILT:MAX?",21,EOI);
<ASCII floating point>
1.106345e-009
• PARAMETER:FILTER:MINIMUM
The PARAMETER:FILTER:MINIMUM command selects the minimum filter time in seconds.
The PARAMETER:FILTER:MINIMUM query returns the minimum filter value.
Command syntax- :CYCLetocycle:PARAMeter:FILTer:MINimum<-2.5 to 2.5>
Example:
Send(0,5,":CYCL:PARAM:FILT:MIN -2.5",25,EOI);
Query syntax- :CYCLetocycle:PARAMeter:FILTer:MINimum?
Example:
Response:
Example:
Send(0,5,":CYCL:PARAM:FILT:MIN?",21,EOI);
<ASCII floating point>
9.941615e-010
• PARAMETER:FUNCTION
The PARAMETER:FUNCTION command selects the current measurement function.
The PARAMETER:FUNCTION query returns the currently selected measurement function.
Command syntax- :CYCLetocycle:PARAMeter:FUNCtion<PW+|PW-|PER+|PER->
Example:
Send(0,5,":CYCL:PARAM:FUNC PER+",22,EOI);
Query syntax- :CYCLetocycle:PARAMeter:FUNCtion?
Example:
Response:
Send(0,5,":CYCL:PARAM:FUNC?",17,EOI);
<PW+|PW-|PER+|PER->
• PARAMETER:SAMPLES
The PARAMETER:SAMPLES command sets the number of Cycle-To-Cycle measurement pairs that are accumulated
each time the ACQUIRE command is issued.
The PARAMETER:SAMPLES query returns the number of Cycle-To-Cycle measurement pairs that are accumulated each
time the ACQUIRE command is issued.
Command syntax- :CYCLetocycle:PARAMeter:SAMPles<1 to 475000>
Example:
Send(0,5,":CYCL:PARAM:SAMP 1000",21,EOI);
Query syntax- :CYCLetocycle:PARAMeter:SAMPles?
Example:
Response:
Example:
Send(0,5,":CYCL:PARAM:SAMP?",17,EOI);
<ASCII integer>
100
138 | Section 6 | Tool Oriented GPIB Commands
©WAVECREST Corporation 2005