Download datasheet: pdf

Transcript
REMOTE OPERATION
b) If the command has more than four letters and the
fourth letter is a vowel, drop it and all letters that follow it. For
instance the command :RESET can be shortened to :RES.
c) If the command has more than four letters and the
fourth letter is a consonant, drop all letters that follow it. For
instance the command :MEASURE can be shortened to :MEAS.
d) If the command contains a question mark (?) or a nonoptional parameter it must be included after the short form
version. For instance a query command of :CONDITION? can be
shortened to :COND?.
e) The use of anything other than the short or long version
of a command is not permitted. For instance both the
:MEASURE and :MEAS commands are acceptable, but :MEASU
is not.
5) If a parameter follows a command it must be separated from
the command by one space.
6) Multiple commands can be sent in one string. The commands
must be separated by semicolons (;). A colon (:) must still
precede each command. If more than one of the commands in
the string requests information from the instrument, the
instrument’s response will also have semicolons separating the
responses, such as 345;0;10.
4-17