Download datasheet: pdf
Transcript
REMOTE OPERATION b) If the command has more than four letters and the fourth letter is a vowel, drop it and all letters that follow it. For instance the command :RESET can be shortened to :RES. c) If the command has more than four letters and the fourth letter is a consonant, drop all letters that follow it. For instance the command :MEASURE can be shortened to :MEAS. d) If the command contains a question mark (?) or a nonoptional parameter it must be included after the short form version. For instance a query command of :CONDITION? can be shortened to :COND?. e) The use of anything other than the short or long version of a command is not permitted. For instance both the :MEASURE and :MEAS commands are acceptable, but :MEASU is not. 5) If a parameter follows a command it must be separated from the command by one space. 6) Multiple commands can be sent in one string. The commands must be separated by semicolons (;). A colon (:) must still precede each command. If more than one of the commands in the string requests information from the instrument, the instrument’s response will also have semicolons separating the responses, such as 345;0;10. 4-17