Transcript
Automatic switch-off Interface Temperature display Time and date format To protect the batteries, the meter switches off automatically when not operated for a longer time. You can select whether switch-off is to take place after one hour or after twelve hours (factory setting: 1 hour). If the data logger is active and during remote interface operation, the auto switch-off feature is disabled. If the Portamess® 913 (X) pH is controlled by a PC and interface conflicts occur when the print key is pressed, you should deactivate the print function (Print OFF) (factory setting: Print On, 4,800 bauds). The transmission speed can be set to 600, 1200, 2400, 4800 or 9600 bauds. The transmission speed must correspond to that set in the printer or PC. Data format and protocol are permanently set to 7 bits, one stop bit, even parity and XON/XOFF protocol (NAMUR NE28). The temperature can be displayed either in °C or °F (factory setting: °C). The time and date format can be set to either 24 hours and day.month.year or 12 hours am/pm and month.day (factory setting: 24 hours and day.month.year). Operation 21