Download User Manual - Laser 2000
Transcript
Quick-Start Tutorial 3.3.8 Average Opens the Profile Averaging menu. Choose either a running average based upon a specific number, Average 5 to 100, of profiles, or choose Continuous (accumulation) averaging of the profiles. An averaging mode indication will appear in blue on the profile graphic. Reset on drift resets the averaging in the centroid position when the drift is (currently) >5 µm. 3.3 The displayed results are derived from the averaged profile. Reset average [ESC]. Press the Esc button to restart the averaging. Enable Average or [CTRL A]. If this is checked (the default), CTRL A toggles averaging on and off. 3.3.9 Filter Filter selects a triangular weighting smoothing function performed as a running smooth of the profile. Select Filter = 0.X% Full Scale. The actual FWHM of the triangular filter may be calculated as: FWHM = [% of full scale x Scale µm/div x Zoom factor]/10 E.g., for: % of full scale = 0.2 Scale µm/div = 50 Zoom factor =4 FWHM = 4 µm Since smoothing causes an increase in the width of the smoothed profile, the displayed value for the beamwidth is corrected using the same algorithm as is used for slit width correction. See Sec. 1.8.3. FWHM Samples BeamMap, Beam’R, ColliMate, BeamScope-P7 3-25