Download User Manual - Laser 2000

Transcript
Introduction
In order to save time for BeamMap users, sections which only apply to Beam’R are indicated by a
double gray line down the left hand side of the page, see left.
DataRay is committed to providing the ultimate in beam profiling performance, and welcomes
constructive criticism of these products and of this manual. Please contact us.
www.dataray.com is always the source for the most current versions of software, manuals,
application notes, specifications, parts lists, etc. At any time, you may download the latest version
of the software and/or manual updates and application notes for free. Visit the website in order to
determine whether an update is available.
If you need a function that is not included in the current version, please fill out and fax a Function
Request form found as Appendix A of this manual. Many requested software functions can be
added with relative ease, and may be done for free, and possibly added to future releases.
However, if the requested embellishment is considered extensive and/or obscure, DataRay
reserves the right to quote a fee for the requested change.
The BeamMap and ColliMate product lines originated in response to OEM requests for production
line instruments to measure focal position. We welcome requests for custom hardware
configurations.
1.2
ABOUT BEAMMAP AND BEAM’R
BeamMap represents a radical new approach to real-time beam profiling. It couples a novel,
patented, multi-plane real-time slit scanning system to improved slit technology, and to the
dramatic increases in processing power represented by continuously evolving PC technology. [A
500 MHz Pentium III is required to fully real-time process the 5 Mega-Samples per second
resulting from eight apertures, with 0.09 µm sample resolution and a 5 Hz update rate.] Initially
designed as a custom product for an OEM customer’s production line, it is now available as a
standard product
If you have every used a single plane beam profiler to measure a tightly focused beam, you will
appreciate the problems that DataRay set out to solve.
Problem 1 - Measuring tightly focused beams.
A focused, aligned in-spec. beam should have the classic
hyperbolic W(z) (diameter) versus z profile shown right. Prior
to alignment and focusing, the beam:
‰ May not be focused at z = 0
‰
May be misaligned in position and angle
‰
May meet neither the beamwaist diameter nor the
irradiance specifications.
‰
May have M2 (Beam Quality) greater than specification
Beam prior to
alignment
Aligned
in-spec.
beam
An instrument with a single plane of measurement gives just one plane of information. With only
a single plane of measurement, finding the focus can be a truly frustrating exercise. Setting
multiple assemblies to the identical focus in a production environment can be costly. If the
assembly will not adjust correctly, you have no real-time indication of what the problem might be.
1-2
BeamMap, Beam’R, ColliMate, BeamScope-P7