Download CO2 Sensor User's Guide
Transcript
4 5 ACINITLO ACHI 6 7 8 ACLO ACONOFF ACPPMHI 9 10 ACPPMLO AMBHI 11 12 AMBLO BCHI 13 200-231 BCLO 8.2.4 occurs after a shorter time. Low byte of above. Autocalibration Interval. Sets the time interval between autocalibrations. Low byte of above. Switches Autocalibration ON/OFF Autocalibration Background Concentration. This determines what background CO2 level is assumed for autocalibration. Low byte of above. Ambient Concentration (for G command). This determines what background CO2 level is assumed for ambient calibration using the “G” command. Low byte of above. Buffer clear time. This will clear any incomplete commands from the serial buffer after a fixed period of inactivity. The time is in half second increments. Low byte of above. User EEPROM 192 94 128 0 1 194 1 194 0 8 255 Autocalibration Settings (locations 3-7) These are included now to maintain compatibility with previous firmware versions. SST recommends using the @ command to set autocalibration timings. 8.2.5 User EEPROM Locations 200 to 231 can be used to store user values. Each location is a single byte. These locations are not used by the sensor. Eg to store the number 42 in the first user EEPROM location: send: response: P 200 42\r\n P 00200 00042\r\n and to read it send: response: p 200\r\n p 00200 00042\r\n Note that the EEPROM is only guaranteed for 100,000 write cycles. © 2013 SST Sensing Ltd AN0114 rev4 21 www.sstsensing.com Please follow all safety information 21