In addition to the header, the measured values shown on the dashboard such as the sample
temperature, magnetic field and slits are logged in the run file.
The following additional commands may be needed in special cases:
END/IMMEDIATE
ABORT
ZFINE
LF
TFX
TFY
TF20
LT
HT
Ends and saves the run without waiting for the period cycle to complete. Use
only if in period mode and the beam has gone off, and you want to take
advantage of the break to change sample.
Ends the run and throws away the data. Use with care!
Selects the Z coil for small longitudinal fields (after F0) or small sweeps around
the main field value (after LF0 and SETMAG to a starting point).
Re-select the main coil after ZFINE. Leaves main field on, turns others off.
Select transverse field in the X direction, no setpoints sent.
Select transverse field in the Y direction, no setpoints sent.
Selects the Y transverse coil and sets 20 Gauss
Put the cryostat in Low temperature mode (manual flow control)
Return the cryostat to the default high temperature mode (auto flow control)
7.5 Temperature Control
The temperature controllers (such as Eurotherm or ITC5) use Proportional – Integral – Differential
(PID) control parameters to adjust the heater power. These are all set from the computer each time
a new setpoint is sent (with SETTEMP from the command line or by a script). Values are read from a
table which can be viewed in SECI under the “Settings” menu.
The first two columns on each line indicate the temperature range over which the parameters listed
are applicable. These should be in increasing temperature with no gaps (if no line matches a new
setpoint, this is an error and it will not be set.) Then follows the maximum power and the P,I,D,
parameters. Negative numbers here will not be set. Finally the required stability, stabilisation time
and timeout limit when running under control of a script.
For experimentation with new PID values you can change the numbers on the temperature
controller’s VI, in one of the tabs under “LabVIEW VIs”. For the ITC5 (Flow and Variox cryostats) you
only need to enter new values (and check that the “PID Source” is set to its usual value of “Labview
Panel”.