Download HiFi User Guide
Transcript
/U Users /C Comments In addition to the header, the measured values shown on the dashboard such as the sample temperature, magnetic field and slits are logged in the run file. The following additional commands may be needed in special cases: END/IMMEDIATE ABORT ZFINE LF TFX TFY TF20 LT HT Ends and saves the run without waiting for the period cycle to complete. Use only if in period mode and the beam has gone off, and you want to take advantage of the break to change sample. Ends the run and throws away the data. Use with care! Selects the Z coil for small longitudinal fields (after F0) or small sweeps around the main field value (after LF0 and SETMAG to a starting point). Re-select the main coil after ZFINE. Leaves main field on, turns others off. Select transverse field in the X direction, no setpoints sent. Select transverse field in the Y direction, no setpoints sent. Selects the Y transverse coil and sets 20 Gauss Put the cryostat in Low temperature mode (manual flow control) Return the cryostat to the default high temperature mode (auto flow control) 7.5 Temperature Control The temperature controllers (such as Eurotherm or ITC5) use Proportional – Integral – Differential (PID) control parameters to adjust the heater power. These are all set from the computer each time a new setpoint is sent (with SETTEMP from the command line or by a script). Values are read from a table which can be viewed in SECI under the “Settings” menu. The first two columns on each line indicate the temperature range over which the parameters listed are applicable. These should be in increasing temperature with no gaps (if no line matches a new setpoint, this is an error and it will not be set.) Then follows the maximum power and the P,I,D, parameters. Negative numbers here will not be set. Finally the required stability, stabilisation time and timeout limit when running under control of a script. For experimentation with new PID values you can change the numbers on the temperature controller’s VI, in one of the tabs under “LabVIEW VIs”. For the ITC5 (Flow and Variox cryostats) you only need to enter new values (and check that the “PID Source” is set to its usual value of “Labview Panel”. 33