With a clean pipette, thoroughly mix the contents of tube 2 and transfer
0.1 ml to tube 3. Discard the pipette; do not mix the contents of tube 3
with the pipette.
Proceed, producing serial, doubling dilutions of the specimen through
tube 8. The prepared serum dilutions are 1:4 to 1:512 for tubes 1 to 8,
respectively. For CSF, the serum dilutions are 1:2 to 1:256. Tube 8 can
be diluted further if an endpoint is not reached.
Test each specimen dilution following the protocol described under
Qualitative Testing.
An independent study was conducted at a clinical laboratory associated with a
large hospital.
Serum and CSF specimens (595) were tested with
Cryptococcus Antigen Test Kit and another commercially available latex test.
The results indicated the sensitivity and specificity of Cryptococcus Antigen
Test Kit is 100% and the reactions were judged easier to interpret.
INTERPRETATION
Qualitative Test:
Positive Result Any agglutination or clumping and clearing of the Test Latex
immediately after the 5-minute rotation step
*Note: Two specimens were negative with Cryptococcus Antigen Test Kit and
positive (titer of 1:2) with the other latex test; both specimens were negative by
culture.
Reproducibility studies were conducted using Cryptococcus Antigen Test Kit
and another commercially available latex test. The majority (96%) of titers
obtained with both tests were within 1 tube dilution of each other
3.
4.
5.
Negative Result -
No agglutination, clumping, or clearing of the Test Latex
immediately after the 5-minute rotation step
Note: The Test Latex suspension should appear milky and
faint traces of granularity may be detectable.
Semi-Quantitative Test:
The titer is the reciprocal of the last dilution which produces a positive result
(agglutination).
EXPECTED VALUES
A positive result indicates the presence of CPS antigen in the patient
specimen. A negative result indicates the absence of detectable levels of CPS
antigen in the specimen.
AIDS patients may retain CPS antigen for long periods of time, which may not
be due to active multiplication of the yeast.10-12 Testing of subsequent
specimens may be useful in determining disease prognosis. Serum antigen
titers ranging from 1:2 to 1:32,678 have been reported in patients with active
HIV infection.13 The magnitude of such titers does not appear to correlate with
the severity of infection but may be useful in diagnosing neurological
cryptococcal infection. Titers >1:10,000 in AIDS patients have been associated
with 100% mortality.
QUALITY CONTROL
All lot numbers of Cryptococcus Antigen Test Kit have been tested and found
to be acceptable. Quality control testing should be performed in accordance
with established laboratory quality control procedures. If aberrant quality
control results are noted, patient results should not be reported.
Quality Control Procedure
The High Positive, Low Positive, and Negative Controls should each be tested
with every test run of patient specimens as described under Qualitative
Testing. The High Positive and Low Positive Controls must agglutinate the
Test Latex differentially.
 The High Positive Control must produce strong agglutination.
 The Low Positive Control must produce weaker but clearly visible
agglutination.
 The Negative Control must not produce any agglutination, although trace
granularity is acceptable.
LIMITATIONS
1. False-positive reactions have been reported with syneresis fluid (agar
surface condensation from culture plates) and with specimens containing
Trichosporon beigelii.14,15
2. Prozone occurs when concentrations of CPS antigens are very high
causing what appears to be a negative result.13 CPS antigen binds the
murine antibody and covers most of the binding sites, preventing
agglutination of the latex. If patient symptomology is reported to be
consistent with cryptococcosis, dilute the CSF or protease-treated serum
specimen 1:10 and repeat testing.
PERFORMANCE CHARACTERISTICS16,17
Sixty randomly selected, known-nonreactive, serum specimens were negative
(nonreactive) when tested with Cryptococcus Antigen Test Kit. Among these
sera were the Centers for Disease Control and Prevention RF reference (1000
IU/ml) samples which included many connective tissue- and rheumatoid factorpositive sera, along with rubella and other nonspecific antibodies. These
results demonstrate Cryptococcus Antigen Test Kit provides optimum
sensitivity without the use of control latex to verify specificity.
Other Latex Test
+
–
Cryptococcus Antigen Test Kit
+
–
68
2*
0
525
BIBLIOGRAPHY
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
Patterson, T.F. and V.T. Andriole. 1989. Eur. J. Clin. Microbiol. Infect. Dis. 8:457-465.
Bloomfield, N., M.A. Gordon, and D.F. Elmendorff, Jr. 1963. Proc. Soc. Exp. Biol.
Med. 114:64-67.
Kaufman, L. and S. Blumer. 1977. Proc. Int. Conf. Mycoses., Pan Am. Health
Org. 356:176-182.
Asznowicz, R., H. Newton-John, and J. Kaldor. 1979. Med. J. Aust. 1:347-349.
Stockman, L. and G.D. Roberts. 1983. J. Clin. Microbiol. 17:945-947.
Gray, L.D. and G.D. Roberts. 1988. J. Clin. Microbiol. 26:2450-2451.
Hamilton, J.R., A. Nobel, D.W. Denning, and D.A. Stevens. 1991. J. Clin.
Microbiol. 29:333-339.
Chosewood, L.C. and D.E. Wilson. 2007. Biosafety in Microbiological and
Biomedical Laboratories. 5th ed. U.S. Dept. of H.H.S., CDC and National
Institutes of Health, Washington, D.C.
Murray, P.R., E.J. Baron, J.H. Jorgensen, M.L. Landry, and M.A. Pfaller. 2007.
Manual of Clinical Microbiology. 9th ed. ASM Press, Washington, D.C.
Kovacs, J.A., A.A. Kovacs, M. Polis, W.C. Wright, V.J. Gill, C.U Tuazon, E.P
Gelmann, H.C. Lane, R. Longfield, G. Overturf, A.M. Macher, A.S. Fauci, J.E.
Parrillo, J.E. Bennett, and H. Masur. 1985. Ann. Int. Med. 103:533-538.
Eng, R.H., E. Bishburg, S.M Smith, and R. Kapila. 1986. Am. J. Med. 81:19-23.
Chuck, S.L. and M.A. Sande. 1989. N. Engl. J. Med. 321:794-798.
Dimech, W.J. 1991. Aust. J. Med. Lab. Sci. 12:13-21.
Boom, W.H., D.J. Piper, K.L. Ruoff, and M.J. Ferraro. 1985. J. Clin.
Microbiol.22:856-857.
Melcher, G.P., K.D. Reed, M.G. Rinaldi, J.W. Lee, P.A. Pizzo, and T.J. Walsh.
1991. J. Clin. Microbiol. 29:192-196.
Howell, D., D. Orkiszewski, and P. Gilligan. 1993. Abstract C-160. Abstracts of
the 93rd Gen. Mtg. of the ASM. ASM, Washington, D.C.
Kiska, D.L., D.R. Orkiszewski, D. Howell, and P.H. Gilligan. 1994. J. Clin.
Microbiol. 32:2309-2311.
PACKAGING
REF R30851501, Cryptococcus Antigen Test Kit............................. 50 Tests/Kit
Symbol Legend
REF
Catalog Number
IVD
In Vitro Diagnostic Medical Device
LAB
For Laboratory Use
Consult Instructions for Use (IFU)
Temperature Limitation (Storage Temp.)
LOT
Batch Code (Lot Number)
Use By (Expiration Date)
EC
REP
European Authorized Representative
IFU 30851501, Revised December 2, 2009
12076 Santa Fe Drive, Lenexa, KS 66215, USA
General Information: (800) 255-6730 Website: www.remel.com Email: remel@remel.com
Local/International Phone: (913) 888-0939 International Fax: (913) 895-4128
Printed in U.S.A