Download CRYPTOCOCCUS ANTIGEN TEST KIT
Transcript
ENGLISH With a clean pipette, thoroughly mix the contents of tube 2 and transfer 0.1 ml to tube 3. Discard the pipette; do not mix the contents of tube 3 with the pipette. Proceed, producing serial, doubling dilutions of the specimen through tube 8. The prepared serum dilutions are 1:4 to 1:512 for tubes 1 to 8, respectively. For CSF, the serum dilutions are 1:2 to 1:256. Tube 8 can be diluted further if an endpoint is not reached. Test each specimen dilution following the protocol described under Qualitative Testing. An independent study was conducted at a clinical laboratory associated with a large hospital. Serum and CSF specimens (595) were tested with Cryptococcus Antigen Test Kit and another commercially available latex test. The results indicated the sensitivity and specificity of Cryptococcus Antigen Test Kit is 100% and the reactions were judged easier to interpret. INTERPRETATION Qualitative Test: Positive Result Any agglutination or clumping and clearing of the Test Latex immediately after the 5-minute rotation step *Note: Two specimens were negative with Cryptococcus Antigen Test Kit and positive (titer of 1:2) with the other latex test; both specimens were negative by culture. Reproducibility studies were conducted using Cryptococcus Antigen Test Kit and another commercially available latex test. The majority (96%) of titers obtained with both tests were within 1 tube dilution of each other 3. 4. 5. Negative Result - No agglutination, clumping, or clearing of the Test Latex immediately after the 5-minute rotation step Note: The Test Latex suspension should appear milky and faint traces of granularity may be detectable. Semi-Quantitative Test: The titer is the reciprocal of the last dilution which produces a positive result (agglutination). EXPECTED VALUES A positive result indicates the presence of CPS antigen in the patient specimen. A negative result indicates the absence of detectable levels of CPS antigen in the specimen. AIDS patients may retain CPS antigen for long periods of time, which may not be due to active multiplication of the yeast.10-12 Testing of subsequent specimens may be useful in determining disease prognosis. Serum antigen titers ranging from 1:2 to 1:32,678 have been reported in patients with active HIV infection.13 The magnitude of such titers does not appear to correlate with the severity of infection but may be useful in diagnosing neurological cryptococcal infection. Titers >1:10,000 in AIDS patients have been associated with 100% mortality. QUALITY CONTROL All lot numbers of Cryptococcus Antigen Test Kit have been tested and found to be acceptable. Quality control testing should be performed in accordance with established laboratory quality control procedures. If aberrant quality control results are noted, patient results should not be reported. Quality Control Procedure The High Positive, Low Positive, and Negative Controls should each be tested with every test run of patient specimens as described under Qualitative Testing. The High Positive and Low Positive Controls must agglutinate the Test Latex differentially. The High Positive Control must produce strong agglutination. The Low Positive Control must produce weaker but clearly visible agglutination. The Negative Control must not produce any agglutination, although trace granularity is acceptable. LIMITATIONS 1. False-positive reactions have been reported with syneresis fluid (agar surface condensation from culture plates) and with specimens containing Trichosporon beigelii.14,15 2. Prozone occurs when concentrations of CPS antigens are very high causing what appears to be a negative result.13 CPS antigen binds the murine antibody and covers most of the binding sites, preventing agglutination of the latex. If patient symptomology is reported to be consistent with cryptococcosis, dilute the CSF or protease-treated serum specimen 1:10 and repeat testing. PERFORMANCE CHARACTERISTICS16,17 Sixty randomly selected, known-nonreactive, serum specimens were negative (nonreactive) when tested with Cryptococcus Antigen Test Kit. Among these sera were the Centers for Disease Control and Prevention RF reference (1000 IU/ml) samples which included many connective tissue- and rheumatoid factorpositive sera, along with rubella and other nonspecific antibodies. These results demonstrate Cryptococcus Antigen Test Kit provides optimum sensitivity without the use of control latex to verify specificity. Other Latex Test + – Cryptococcus Antigen Test Kit + – 68 2* 0 525 BIBLIOGRAPHY 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. Patterson, T.F. and V.T. Andriole. 1989. Eur. J. Clin. Microbiol. Infect. Dis. 8:457-465. Bloomfield, N., M.A. Gordon, and D.F. Elmendorff, Jr. 1963. Proc. Soc. Exp. Biol. Med. 114:64-67. Kaufman, L. and S. Blumer. 1977. Proc. Int. Conf. Mycoses., Pan Am. Health Org. 356:176-182. Asznowicz, R., H. Newton-John, and J. Kaldor. 1979. Med. J. Aust. 1:347-349. Stockman, L. and G.D. Roberts. 1983. J. Clin. Microbiol. 17:945-947. Gray, L.D. and G.D. Roberts. 1988. J. Clin. Microbiol. 26:2450-2451. Hamilton, J.R., A. Nobel, D.W. Denning, and D.A. Stevens. 1991. J. Clin. Microbiol. 29:333-339. Chosewood, L.C. and D.E. Wilson. 2007. Biosafety in Microbiological and Biomedical Laboratories. 5th ed. U.S. Dept. of H.H.S., CDC and National Institutes of Health, Washington, D.C. Murray, P.R., E.J. Baron, J.H. Jorgensen, M.L. Landry, and M.A. Pfaller. 2007. Manual of Clinical Microbiology. 9th ed. ASM Press, Washington, D.C. Kovacs, J.A., A.A. Kovacs, M. Polis, W.C. Wright, V.J. Gill, C.U Tuazon, E.P Gelmann, H.C. Lane, R. Longfield, G. Overturf, A.M. Macher, A.S. Fauci, J.E. Parrillo, J.E. Bennett, and H. Masur. 1985. Ann. Int. Med. 103:533-538. Eng, R.H., E. Bishburg, S.M Smith, and R. Kapila. 1986. Am. J. Med. 81:19-23. Chuck, S.L. and M.A. Sande. 1989. N. Engl. J. Med. 321:794-798. Dimech, W.J. 1991. Aust. J. Med. Lab. Sci. 12:13-21. Boom, W.H., D.J. Piper, K.L. Ruoff, and M.J. Ferraro. 1985. J. Clin. Microbiol.22:856-857. Melcher, G.P., K.D. Reed, M.G. Rinaldi, J.W. Lee, P.A. Pizzo, and T.J. Walsh. 1991. J. Clin. Microbiol. 29:192-196. Howell, D., D. Orkiszewski, and P. Gilligan. 1993. Abstract C-160. Abstracts of the 93rd Gen. Mtg. of the ASM. ASM, Washington, D.C. Kiska, D.L., D.R. Orkiszewski, D. Howell, and P.H. Gilligan. 1994. J. Clin. Microbiol. 32:2309-2311. PACKAGING REF R30851501, Cryptococcus Antigen Test Kit............................. 50 Tests/Kit Symbol Legend REF Catalog Number IVD In Vitro Diagnostic Medical Device LAB For Laboratory Use Consult Instructions for Use (IFU) Temperature Limitation (Storage Temp.) LOT Batch Code (Lot Number) Use By (Expiration Date) EC REP European Authorized Representative IFU 30851501, Revised December 2, 2009 12076 Santa Fe Drive, Lenexa, KS 66215, USA General Information: (800) 255-6730 Website: www.remel.com Email: remel@remel.com Local/International Phone: (913) 888-0939 International Fax: (913) 895-4128 Printed in U.S.A