General Technical Specifications
Medical CE-mark:
The CE-mark indicates that Interacoustics A/S meets the requirements of Annex II of
the Medical Device Directive 93/42/EEC.
Approval of the quality system is made by TÜV - identification no. 0123.
IEC 60601-1, Class I, Type B
(using UPS400 external power supply)
IEC 60601-1-2
EMC:
IEC 60645-1/ANSI S3.6 , Type 3 Tone audiometer
Audiometer:
UPS400 External power supply
Power:
See description in separate Service Manual
2xT1A + 1xT2A located on PCB
Fuses:
Operation environment: Temperature:
15 – 35 C
30 – 90%
Rel. Humidity:
Transport temperature:: -20-50°C
0-50°C
Storage temperature:
Warm up time:
3 minutes at room temperature (20 C).
Standards:
Safety:
Functions
Signal sources:
Tone:
Masking:
Inputs:
Outputs:
Pat. response
Power UPS400
Earphone Right /
Left:
Bone conductor:
Insert masking
phone
MTP10:
Printer:
RS232
11 pure tone and warble test frequencies in the range of
125 – 8000Hz
Narrow band noise, in 11 bands.
White noise.
Connection for patient response switch. Response lamp will
light when activated.
Connection for external power supply.
Can be selected to be either standard earphones or insert
phones by pressing shift Right/Left.
Masking in Right or Left earphone can be selected.
For use of CIR22.
Use of MTP10 as printer device
Laser printer with HP GL2 language
Ink jet or matrix printer using IBM mode (optional)
Desk jet printers using HP PCL L3 language (optional)
Input/output for computer communication. See separate
section in Service Manual for programming details
Operator to patient communication. 10 – 110 dB
continuously adjustable on the operation panel
Patient
Communication:
Talk Forward:
Attenuator:
-10 to 120 dB in 1 or 5 dB steps for tone and speech. The upper 20 dB is accessed
by pressing the +20 dB button.
0 – 110 dB in 1 or 5 dB steps for masking noise.
Silent touch switch for soft presentation of signals. Tone lamp will light when
activated.
Tone is presented by activation of Tone Switch.
Manual:
Tone is interrupted by activation of Tone Switch.
Reverse:
Single pulse or automatic pulsing with selectable on/off times of
Pulse:
250 – 5000ms.
Patient controlled Hughson-Westlake test according to ISO 8253-1 with automatic
storing of the threshold values in the internal memory.
Patient response is open for 2 seconds after signal presentation.
2 out of 3 or 3 out of 5 correct responses can be selected.
Pure tone or narrow band stimulation. Fixed frequency. Continuous and pulsed tone.
If selected, it will indicate end of automatic test.
With automatic score calculation (5 dB included for
SISI:
familiarisation).
Binaural pure tone or speech stimulation.
Stenger:
Tone Switch:
Tone presentation:
Auto Threshold:
Békésy:
Buzzer:
Special test types:
AD226 Service Manual 2