Download Avance Anesthesia Machine
Transcript
Avance 2.9 Electronic Gas Mixer The Gas Mixer receives its pneumatic inputs from the pipeline and cylinder supplies and sends mixed gas to the vaporizer manifold. The Gas Mixer interfaces to the Anesthesia Control board for power and communications. The Gas Mixer consists of the following subassemblies and main components: • Gas Mixer board (A) • Control Manifold (B) — manifold, selector valves, proportional valves • Flow sensor assembly (C) • Mixed gas manifold and exit check valve (D) A D C B Figure 2-21 • Electronic Gas Mixer 2-20 09/07 1009-0357-000