Download Aespire 7900 Anesthesia Machine Technical Reference Manual
Transcript
9 Repair Procedures 9.22.3 Alarm speaker To replace the alarm speaker: 1. To access the speaker, follow the disassembly instructions in Section 9.22.1. 2. Remove the two screws holding the alarm speaker to the housing. 3. Install the new alarm speaker using the previously removed screws. 4. Assemble in reverse order. 5. Perform the Checkout Procedure found in Section 3. 1009-0541-000 02/05 9-53