Section 4 – Troubleshooting Table
Problem
Cause
Air passing through meter.
Over-registration (registering
more
volume
than
pass
Calibrator out of adjustment.
ing through meter).
Leaky valve in prover system (during
test run).
Under-registration (register- Mechanical friction in meter.
ing less volume than passing through meter).
Binding or misalignment in driven
accessories.
Loose gear train or slipping drives.
Calibrator out of adjustment.
Oversize clearances in meter.
Product being eliminated through
vent line.
Product being eliminated
Air eliminator cylinder seal damaged.
through vent line.
Air eliminator float ruptured.
Foreign particles on cylinder sealing
surfaces.
System fails to shut off
Linkage out of adjustment between
set-stop counter and valve.
properly after delivering
preset
quantity.
Improper setting of “Minimum Setting”
in set-stop counter.
Mechanical obstruction in valve.
Defective pushbutton set-stop
counter.
Plugged orifice.
Strainer clogged.
Restricted product flow
through
meter.
Butterfly valve in air eliminator
restricting flow.
Back pressure or manual valves do
not open.
Mechanical obstruction in meter.
Pump not operating (pressure
system).
No product.
Air being eliminated from product.
Pump undersized.
Pump not operating properly.
System leaks after
shutdown. Linkage out of adjustment.
Defective main seat seal in valve.
Foreign matter holding main valve
seat open.
Drive coupling defective.
Large numeral counter or
ticket printer not operating
Internal problems.
satisfactorily.
Page 18 • MN01029
Remedy
Check air eliminator.
Readjust calibrator. See MN01028, Pg. 4.
Repair prover valve and rerun test.
Check all meter clearances. See Pgs. 12
through 13. Check bearings for tightness.
Inspect accessories. Repair or replace.
Repair or replace.
Readjust calibrator. See MN01028, Pg. 4.
Check meter clearances. See Pg. 12.
See below.
Replace cylinder seal.
Replace float.
Clean sealing surfaces.
Adjust linkage correctly.
See Pushbutton Set-Stop Counter
Service Manual.
Disassemble and remove obstruction.
Inspect valve internals for damage.
See Troubleshooting Section in Pushbutton Set-Stop Counter Service Manual.
Check orifice in upper piston of P-2 or
PG-2 Valve (pressure system only).
Clean strainer. Replace if necessary.
Repair or replace.
Disassemble and inspect for foreign
particles and/or defects. Repair.
Disassemble and remove obstruction.
Inspect meter internals for damage.
Inspect pump for mechanical or electrical defects. Replace if necessary.
Supply product.
Normal - product flow automatically is
restricted during air elimination.
Replace with pump of sufficient capacity.
Check pump.
Adjust linkage correctly.
Replace seal.
Disassemble valve and remove foreign
matter. Check seal for damage.
Replace.
See Veeder Root Service Manual.