MV2000(Touch)
Ver 2.0
tBILEVEL Mandatory Breath
MEKANT’s tBILEVEL mode is complex mode it may happened Mandatory Breath and
Spontaneous Breath. Because it works as SIMV mode but it difference because
Spontaneous range not fixed. So, anytime it is able to Spontaneous Breath.
tBILEVEL has two level of PEEP to supply patient. The upper PEEP is PEEPHIGH and
lower PEEP is PEEPLOW. Mandatory Breath’s Inspiration is from PEEPLOW to PEEPHIGH
and Expiration is from PEEPHIGH to PEEPLOW . So tBILEVEL mode’s Mandatory Breath
is Pressure Control Ventilation. Each change time of before and after small zone is
changing zone, this zone purpose is to detect patient breath if it detect patient breath it
would adjust little early or late to start change inspiration and expiration for
synchronizing with patient breathing.
Setting parameters are Respiratory Rate, PEEPHIGH, PEEPLOW and so on, In case of
Pressure Support Ventilation there is more add Support Pressure, FAP, Essens and so
on.
MEKANT’s tBILEVEL mode supplies 1.5 cmH2O Pressure in Spontaneous Breath. If
user want to more high pressure support it will use Pressure Support Ventilation setting.
If Support Pressure + PEEPL value is higher than PEEPH It will be working pressure
support ventilation difference than Spontaneous breath works Pressure Support
Ventilation. Refer to the figure below.
64