Download preliminary
Transcript
P R E L I M I N A R Y GE HEALTHCARE PART NUMBER FN091019 REVISION 2 8-6-2 VS5 N & VS6 N SERVICE MANUAL Battery Replacement Procedure (cont’d) 5) Remove the metal securing bracket (with the screw attached) by sliding it to the side and lifting it up. Battery Securing Bracket Figure 8-134 Removing the Battery Securing Bracket 6) Slide the Battery upwards and out of the battery compartment along the grooves - until it is fully removed from the system (see Figure 8-135). Figure 8-135 Sliding the Vivid S6 N Battery out of the Battery Compartment NOTE: WHEN REMOVING A DEFECTIVE BATTERY, ENSURE THAT IT IS DISPOSED OF IN ACCORDANCE WITH LOCAL REGULATIONS. ALTERNATIVELY, FORWARD IT TO GE HEALTHCARE FOR PROPER DISPOSAL. Chapter 8 - Replacement Procedures 8-107