VS5 N & VS6 N SERVICE MANUAL
Battery Replacement Procedure (cont’d)
5) Remove the metal securing bracket (with the screw attached) by sliding it to the side and lifting it up.
Battery Securing Bracket
Figure 8-134 Removing the Battery Securing Bracket
6) Slide the Battery upwards and out of the battery compartment along the grooves - until it is fully
removed from the system (see Figure 8-135).
Figure 8-135 Sliding the Vivid S6 N Battery out of the Battery Compartment
NOTE:
WHEN REMOVING A DEFECTIVE BATTERY, ENSURE THAT IT IS DISPOSED OF
IN ACCORDANCE WITH LOCAL REGULATIONS. ALTERNATIVELY, FORWARD IT
TO GE HEALTHCARE FOR PROPER DISPOSAL.