Performance Verification
Scope Input A, B, C, D Tests
4
Table 4-11. Input B Trigger Sensitivity Test Points
UUT
UUT
Model
Time base
190-502
5502A SC...
Initial Input Voltage
MODE levsin
UUT
Trigger Amplitude
Frequency
200 ns/div
100 mV pp
5 MHz
0.5 div
10 ns/div
400 mV pp
60 MHz
1 div
10 ns/div
800 mV pp
100 MHz
2 div
190-102
10 ns/div
400 mV pp
100 MHz
1 div
190-104
10 ns/div
800 mV pp
150 MHz
2 div
190-202
10 ns/div
400 mV pp
200 MHz
1 div
190-204
10 ns/div
800 mV pp
250 MHz
2 div
190-502
2 ns/div
400 mV pp
500 MHz
1 div
190-504
2 ns/div
800 mV pp
600 MHz
2 div
190-062
Input B AC Voltage Accuracy (HF) and Bandwidth Test
Proceed as follows to test the Input B high frequency automatic scope ac voltage
measurement accuracy and the bandwidth:
1. Connect the Test Tool to the 5502A as shown in Figure 4-8.
2. Select the Test Tool setup:
SAVE
a) Recall the setup (see Standard Test Tool Setup). Press
, F2
ENTER
(RECALL) and select SETUP, press
, select the setup name, and
ENTER
press
to recall the setup.
b) Press SCOPE , then press F2
(READING…), select
(READINGS 2), and select on B | V ac.
c) Press
MANUAL
AUTO
mV
F1
to select autoranging (AUTO in upper right LCD edge).
RANGE
d) Use RANGE and V to change the sensitivity range to select manual
sensitivity ranging and lock the Input B sensitivity range on 500 mV/div.
e) Press TRIGGER and use F1
to select Input B as trigger source.
3. Set the 5502A to source a sine wave at the first test point in Table 4-12.
4. Observe the Input B reading and check that it is within the range shown
under the appropriate column of Table 4-12.
5. Continue through the test points.
4-25