Download Service Manual
Transcript
Calibration and Verification Verification 3 2. Using the N to BNC cable, connect the CHAN 1 connector on the Calibrator Mainframe to the Fluke 5520A Calibrator voltage binding posts using a BNC to banana connector. Record the 5820A reading for each voltage in Table 3-10. Table 3-10. DC Voltage Measurement Verification Nominal Value (V dc) Measured Value Deviation 1-year Spec. (V p-p) 0V 0.001 1V 0.0015 2.5 V 0.00225 5V 0.0035 6.5 V 0.02625 9.9 V 0.03475 -1 V 0.0015 -2.5 V 0.00225 -5 V 0.0035 -6.5 V 0.02625 -9.9 V 0.03475 3-33. Current Verification This procedure uses the following equipment: • Hewlett-Packard 3458A Digital Multimeter with cable • 1 Ω, 10 Ω, 100 Ω, and 1000 Ω precision resistor assemblies Remove the front panel Current Loop using the 2 mm hex head driver. 3-34. DC Current Verification 1. Connect the DMM directly to the vacant banana jacks on the calibrator. 2. Program the Calibrator Mainframe to output the Currents listed in Table 3-11. Press on the Calibrator Mainframe to activate the output. 3. Allow the HP 3458A reading to stabilize, then record the HP 3458A reading for each current in Table 3-11. 4. Compare result to the tolerance column. 3-27