2-20 Calibration Manufacturing calibration The manufacturing calibration procedure is normally performed only at the factory, but the necessary steps are included here in case the unit is repaired, and the unit requires these calibration procedures. NOTE If the unit has been repaired, the entire comprehensive calibration procedure should also be performed in addition to the manufacturing calibration procedure. Firmware revision level The manufacturing calibration procedure depends on the firmware revision level, either revision A13 and lower, or revision A14 and higher. The instrument displays the firmware revision as part of the power-up cycle. Be sure to use the correct procedure as outlined below. Recommended test equipment Table 2-13 summarizes the test equipment required for the manufacturing calibration steps. Note that the equipment depends on firmware revision level. See Table 2-1 for equipment necessary for the comprehensive calibration steps. Table 2-13 Recommended equipment for manufacturing calibration Firmware revision A13 and lower Keithley 3930A or 3940 Frequency Synthesizer: 1V RMS, 3Hz, ±5ppm 1V RMS, 1kHz, ±5ppm Keithley Model 2001 or 2002 Digital Multimeter: 1V, 3Hz AC, ±0.13% Keithley Model 8610 Low-thermal short Firmware revision A14 and higher Stanford Research Systems DS 345 Function Generator: 1V RMS, 1kHz, ±5ppm Fluke 5700A Calibrator: 1V RMS, 10Hz, ±260ppm Keithley Model 8610 Low-thermal short