Download T940/T964 User Manual - Astronics Test Systems

Transcript
Model T940 User Manual
Publication No. 980938 Rev. K
The relevant VXIplug&play API function is:
•
tat964_setVoltageRangeMode
Note: This function is inoperable for the DR1, DR2, DR4, DR7, and DR8.
MFSIG Source
This pull-down control programs the power connector MFSIG signal function.
The DR3e Driver/Receiver boards have an optional front panel power connector
that is used to provide the V+/V- rail voltages to the Pin Electronics devices.
In addition, a signal is provided that can be programmed to generate a shutdown
level to the external voltage source or to light an LED. The selections for this
pull-down control are:
Table 5-17: MFSIG Settings
Setting
Shutdown High
Shutdown Low
Disabled
MPSIG
Description
Signal goes high on voltage or temperature fault
condition
Signal goes low on voltage or temperature fault
condition
Signal is not driven
Signal is assigned to the sequencer MPSIG signal
The relevant VXIplug&play API function is:
•
tat964_setPowerSettings
MPSIG Signal
This control sets the source of the MPSIG. All checked signals are ORed
together.
Table 5-18: MPSIG Source
Setting
Sequence Active
Description
MPSIG goes high when sequence active is true.
Paused
MPSIG goes high when the sequencer is paused.
Halt
MPSIG goes high when the sequencer is halted.
Burst Error
Over Current
Drive Fault
MPSIG goes high when burst error is true.
MPSIG goes high when over current is true.
MPSIG goes high when drive fault is true.
Watchdog Timeout
MPSIG goes high when the watchdog timeout is true.
Sequence Timeout
MPSIG goes high when the sequence timeout is true.
Pattern Timeout
Sync Error
Soft Front Panel Operation 5-22
MPSIG goes high when the pattern timeout is true.
MPSIG goes high when the sync error is true.
Astronics Test Systems