Download 8.2 Optical unit

Transcript
11002644_AA.book Page 102 Monday, February 21, 2005 4:45 PM
8
Installing and modifying components
8.2.5
Changing the lamp assembly (optional)
WARNING! The system uses high intensity ultra-violet light. Do not remove
the UV lamp while the system is running. Before replacing a UV lamp, ensure
that the lamp cable is disconnected from the rear of the system to prevent
injury to the eyes. If the mercury lamp is broken, make sure that all mercury is
removed and disposed according to national and local environmental
regulations.
Lamp housing end plate
1
Use a screwdriver to detach the end plate by removing one and loosening
the other of the two holding screws on the lamp housing to be removed.
2
Slide the lamp housing off the filter housing.
3
Detach the end plate, as in step 1 above, from the lamp housing to be fitted
to the optical unit.
4
Slide the lamp housing onto the filter housing. The lamp and signal cables
should be on the same side. As you slide the lamp housing into position,
depress the two pressure pads on the filter housing in sequence to facilitate
the installation.
Dot on lamp housing
Symbols on
filter housing 5
Symbol on
filter wheel
below lid
102
Refit the lamp housing end plate.
6
Slide the lamp housing firmly into place. There will be a faint click when the
housing is positioned correctly. The Hg lamp housing can take up two
positions, one for 280 nm, marked by
on the filter housing, and the other
marked by for all other wavelengths. The Zn lamp housing has only one
position.
7
Set the wavelength to be used by selecting lamp position (indicated by a dot
on the lamp housing) in combination with the appropriate filter, i.e. the dot on
the lamp housing should be adjacent to the symbol on the filter housing
corresponding to the symbol on the filter wheel for the filter to be used. A
click will indicate that the filter is in position.
ÄKTAprime plus User Manual 11-0026-44 Edition AA