Installing and modifying components
8.2.3
Connecting the optical unit to the system
If the optical unit has been disconnected from the system, connect it as follows:
1
Place the optical unit in the holder.
2
Secure it by tightening the screw in the
holder.
3
Connect the lamp cable to the socket UV
Lamp on the rear panel of the module.
4
Holder
Optical unit
Connect the signal cable to the socket UV on the rear panel of the module.
8.2.4
1
Screw
Connection to the column
Fix the optical unit directly under the column if possible.
Note: Always position the optical unit with the filter wheel cover facing
upwards.
2
Connect the column outlet tubing directly onto the top of
the optical unit using a fingertight connector and screw to
finger-tightness.
3
Connect the optical unit outlet tubing onto the opposite
hole in the flow cell. Use fingertight connectors.
If no outlet tubing exists, cut a piece of PEEK tubing (i.d.
0.75 mm, o.d. 1/16"). The length should be 170 mm.
4
Connect the other end of the tubing to the conductivity flow cell.
ÄKTAprime plus User Manual 11-0026-44 Edition AA