If desired, the dispensation order can be entered manually. Note: When clicking “Generate Dispensation Order”, any existing dispensation order will be overwritten. Note: When a base position is selected in the sequence to analyze, the corresponding dispensation is highlighted with a gray background color, and vice versa. Dispensation warnings If the dispensation order contains a warning, this is displayed by a red exclamation mark at the end of the text box. It is possible to run an assay with a dispensation warning, but the warning must be considered when evaluating the analysis result. If you position the mouse pointer over the exclamation mark, a tooltip will display an explanation of the warning. Warning Suggested action Sequence uncertain due The problem may be resolved by either entering to lack of terminal more sequence information or reducing the sequence information. number of dispensations. Sequence not in phase at the end of the dispensations. The problem may be resolved by adjusting the dispensation order (manually or by clicking “Generate Dispensation Order”) or entering more sequence information. Note: If the problem is not resolved, the out-of phase stretch will not be analyzed. The generated dispensation order contains less reference peaks than required. 16 If possible, enter more sequence information and increase the number of dispensations. For the best possible quality assessment of the results, we recommend five or more reference peaks. PyroMark Q24 Analysis Software User Guide 01/2009