the trigger polarity setting, i.e., bit 5 of the channel control and status
register A (ChanCSRA). For very low and very big DAC values the
ADC will be driven out of range and an unpredictable, but constant
response is seen. From the sloped parts a user program can find the
DAC value that is necessary for a desired ADC offset. It is
recommended, that for unipolar signals an ADC offset of 1638 units is
chosen. For bipolar signals, like the induced waveforms from a
segmented detector, the ADC offset would be 8192 units, i.e., midway
between 0 and 16384.
A user program would use the result from the calibration task to find,
set and program the correct offset DAC values.
Since the offset measurement has to take the preamplifier offset into
account, this measurement must be made with the preamplifier
connected to the PIXIE-4 input. The control task makes 16
measurements at each DAC step and uses the last computed DC-offset
value to enter into the data buffer. Due to electronic noise, it may
occasionally happen that none of the sixteen attempts at a base line
measurement is successful, in which case a zero is returned. The user
software must be able to cope with an occasional deviation from the
expected straight line.
On exit, the task restores the offset DAC values to the values they had
on entry.
ControlTask 4:
Untriggered Traces
This task provides ADC values measured on all four channels and
gives the user an idea of what the noise and the DC-levels in the
system are. This function samples 8192 ADC words for the channel
specified in CHANNUM. The XWAIT variable determines the time
between successive ADC samples (samples are XWAIT * 13.3ns
apart). In the PIXIE-4 Viewer XWAIT can be adjusted through the dT
variable in the Oscilloscope panel. The results are written to the 8192
words long I/O buffer. Use this function to check if the offset
adjustment was successful.
From the PIXIE-4 Viewer this function is available through the
Oscilloscope Panel. Hit the Refresh button to start four consecutive
runs with ControlTask 4 in the selected module, one for each channel.
ControlTask 5:
ProgramFiPPI
This task writes all relevant data to the FiPPI control registers.
ControlTask 6:
Measure Baselines
This routine is used to collect baseline values. Currently, DSP collects
49
PIXIE-4 Programmer’s Manual V1.41