Download Digital Gamma Finder (DGF)
Transcript
a) read from address 0x0 128K words (= 512K bytes) to read all spectra, b) read from address 0x2000 two times the number of words specified in DSP variables EMwords (= number of 16 bit list mode data words acquired), i.e. ( 2 x EMwords ) bytes c) in double buffer mode, read from address 0x20000 [or 0x30000] two times the number of words specified in DSP variables EMwords [or EMwords2] d) read from address 0x10000000 2 bytes to obtain the serial number. The USB chip has an internal FIFO for 512 bytes. It is therefore most efficient to read data in multiples of 512 bytes; reading fewer bytes takes about the same time (or more?) that 512. 23 DGF-4C – Pixie-4 Differences V1.2 © XIA 2009. All rights reserved.