User manual Endurance
Peak
number
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
Area
206342
204773
205425
205203
205566
205568
206156
206174
205956
206165
205765
205960
206296
205617
205619
205648
206147
207070
From the integration results the RSD% for the Endurance
can be calculated.
For this example the result is:
RSD% = 0.25 %
The specified value is an RSD% < 1%.
E.1.4 Mixing and carry-over test
The default system settings were used, except for the following:
<USAGE>
Use of mix methods: ENABLED
A plate with one sample well filled with sample (50 ppm Uracil, A01), one sample well filled
with eluent (A02), one empty (destination, A03) well and one Reagent_A vial are placed in
the Endurance. A 1:10 dilution of sample from the sample well will be made in the empty
destination well. After dilution from both wells (sample and destination) 10 µL will be injected.
The Reagent-A vial is filled with eluent.
The wash solvent bottle is filled with 80% H2O / 20% Iso-propanol.
0060.183.21