D 500
The pump is running and then is showing
a warning or fault message
Incident
Warning
W16 ADP TEMP. LOW
Pump temperature is < of 15°C to the
temperaturecontrol setting value
Warning
Cause
Temperature control valve supply
problem
Troubleshooting
y
Other problem
Incorrect monitoring configuration
Check the temperature control valve
connection on Therm. ADP connector
(see D 10).
Call customer service center.
y
W18 LOAD-LOCK PUMP
Check that “LL PUMP OPTION” is not
valid. (see C 42).
LI1 and LI3 logic inputs are not used
to monitor a LL pump
LL pump problem
W18 LOAD-LOCK PUMP
Wiring problem
LI1 and LI3 logic inputs are used to
monitor a LL pump
Other problem
Call customer service center.
y
Check the connections on LI1 and LI3
connectors (see D 10).
Call customer service center.
The maintenance threshold of a
parameter is reached
y
Check maintenance thresholds.
Call customer service center.
The maintenance threshold of a
parameter is reached
y
Check maintenance thresholds.
Call customer service center.
The maintenance threshold of a
parameter is reached
y
Check maintenance thresholds.
Call customer service center.
Warning
Warning
W31 ADP LP MAINT.
The maintenance threshold of a
parameter is reached
y
Check maintenance thresholds.
Call customer service center.
Warning
The maintenance threshold of a
parameter is reached
y
Check maintenance thresholds.
Call customer service center.
Incorrect M4 configuration
y
Configure for “ADP 122“ or “ADS“
(see C 42).
Warning
W28 LOAD-LOCK MAINT.
Warning
W29 ROOTS MAINT.
Warning
W30 EXHAUST MAINT.
W32 ADP MAINT.
Fault on ADS 1202/1802
D34 AR.URGT; CONT.ADP
Fault
D35 EMERGENCY STOP
6/7
Other problem
Pump emergency stop is engaged
Other problem
Call customer service center.
y
Rotate to unlock the emergency stop
on the front panel display.
Call customer service center.
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two
G 03219 - Edition 01 - December 06