Download C 500
Transcript
C 500 Basic checks p Check the water differential pressure The water differential pressure between water inlet and outlet must be at least 3 bar. Water IN flow meter Water OUT flow meter If not: Utility control panel check the delivery water system check that the water inlet and outlet valves are open, and if not, open them: they must be in the same position as shown on the photo. check the water circuit. Water outlet valve Water inlet valve Utility connection panel ams01402.doc - Edition 02 – May, 06 q Check the He supply pressure The He supply pressure must be 1.5 bar + 0.3 bar. If reading is less than 1.5 bar, check the He supply line. Alcatel Vacuum Technology France – AMS 200 User’s manual 3/6