Troubleshooting E. TROUBLESHOOTING Possible procedural and system problems, as well as recommended solutions. If you cannot resolve the problem, call technical service at 800-345-0206. No power to the pipette. Probable Cause Recommended Solution Batteries are incorrectly installed. Recheck the orientation of the batteries; they should match the diagram in the battery case Battery pack cables are not properly connected. Ensure the connections to the battery pack are securely plugged into the connection from the pipette body. Battery Spacer is not in unit or is not in correctly (Multichannels only with NiMH). Ensure that the battery spacer is in place for all models that use Nickel Metal Hydride (see section F.3.1). Double beep tone occurs. Probable Cause Recommended Solution An illegal programming step is being entered, such as programming two fills that exceed the maximum fill volume. Recheck the programming sequence. Ensure you are not attempting to fill above the maximum fill volume of the pipette, or attempting to dispense more than the total fill volume. “ERR_1” is displayed. (Signifies one of the following conditions) Probable Cause Recommended Solution Liquid is aspirated into the pipette. Disassemble and clean the pipette. (See Section F.1 and/or F.2, for pipette cleaning and autoclaving instructions). Motor is defective or the flex circuit assembly is damaged. Contact technical service at: 800-3450206 29