Troubleshooting
E. TROUBLESHOOTING
Possible
procedural
and
system
problems,
as
well
as
recommended
solutions.
If
you
cannot
resolve
the
problem,
call
technical
service
at
800-345-0206.
No power to the pipette.
Probable
Cause
Recommended
Solution
Batteries are incorrectly
installed.
Recheck
the
orientation
of
the
batteries;
they
should
match
the
diagram
in
the
battery
case
Battery
pack
cables
are
not
properly
connected.
Ensure
the
connections
to
the
battery
pack
are
securely
plugged
into
the
connection
from
the
pipette
body.
Battery
Spacer
is
not
in
unit or is not in correctly
(Multichannels only with NiMH).
Ensure
that
the
battery
spacer
is
in
place
for
all
models
that
use
Nickel
Metal
Hydride
(see section F.3.1).
Double beep tone
occurs.
Probable
Cause
Recommended
Solution
An
illegal
programming
step
is
being
entered,
such
as
programming
two
fills
that
exceed
the
maximum
fill
volume.
Recheck
the
programming
sequence.
Ensure
you
are
not
attempting
to
fill
above
the
maximum
fill
volume
of
the
pipette,
or
attempting
to
dispense
more
than
the
total
fill
volume.
“ERR_1” is displayed. (Signifies
one
of
the
following
conditions)
Probable
Cause
Recommended
Solution
Liquid
is
aspirated
into
the
pipette.
Disassemble
and
clean
the
pipette.
(See
Section F.1 and/or F.2, for pipette cleaning
and autoclaving instructions).
Motor
is
defective
or
the
flex
circuit
assembly
is
damaged.
Contact
technical
service
at:
800-3450206
29