Download Matrix Electronic Pipettes
Transcript
Maintenance 3. Disconnect wire connectors. Pull apart the two wire connectors over the battery pack (see Figure E.3.1Ab). 4. Remove battery pack. Pull gently on the battery wire connector (without the pins) to dislodge the battery pack, then lift the battery pack out of the main housing (see Figure F.3.1Ac). Figure F.3.1Ab: Disconnect Wires 5. Insert new battery pack. Insert the new battery pack into the cavity of the main housing. Ensure that the wire is looped under the battery pack and the connector is above it (refer to Figure F.3.1Ac). NOTE: The battery pack fits very snugly in the main housing; therefore, ensure that the pack is oriented as shown in Figure F.3.1Ac. The connectors for the battery pack and the pipette should be on opposite sides. Figure F.3.1Ac: Routing battery wires 6. Reconnect wire connectors. Reconnect the wire connectors above the battery pack. Gently push the wires of the battery pack down until the wire connectors lay flat over the top of the battery pack. Figure F.3.1Ad: Reconnect Wires 45