Chapter 7 Scope Starting up with the scope Starting up with the scope Once the FiberScope icon is validated: 1 Press RESULTS hard key Focus level for image sharpness Fig. 26 Example of the result using the P5000i scope Use the Focus Control button onto the P5000i scope (see Figure 22 on page 48) to adjust the image quality and sharpness. NOTE To switch from Scope page to FO results page and vice-versa, press the RESULTS hard key for about 2 seconds (a beep is emitted). Freeze mode Once the image is acceptable, you may freeze the picture. This feature allows to store in memory the resulting picture. Freezing a scope result does not store the picture in a file (see "File menu" page 58). The result will be lost if the instrument is shut off, or if more than 3 pictures are frozen (see "Mosaic Mode" page 56). NOTE The button set on the lead, or the QuickCapture™ on the P5000 allows to freeze the picture, and to launch a test (according to configuration in Setup page - see “Configuring the Scope” on page 49). User Manual 760000002/10 53