ii
SPECIFICATIONS
MECHANICAL
Size: 12"L x 4.5"W x 4.5"H excluding handle, handle extends height to 6.25".
Weight: 5 lbs, 2 oz including alkaline batteries.
External Controls: 5 pushbutton switches for POWER ON/OFF, MODE, ALTERNATE, RESET and LIGHT.
Internal Controls: Cal and Reset pushbutton;Coarse Contrast, Pole Zero, High
Voltage, and LLD trimpots.
DETECTOR
Type: Sealed Spherical TE Proportional Counter. Rossi Type. 2 1/4" ID.
Wall Material: A150 Conducting Tissue Equivalent plastic.
Wall Thickness: 0.12 cm, 144 mg/cm, Aluminum can .065"
Filling Gas: Propane gas, 2 micron
Internal Source: Less than 1 uCi Cm244
OPERATIONAL
Readout: Alphanumeric 2 line x 16 character LCD
Range: Autoranging from .001 mREM/h to 999 REM/h and .001 mRAD/h to 999
RAD/h rate and .001 mREM to 999 REM and .001 mRAD to 999 RAD integrate.
SI units of Sv and Gy are also selectable.
Multi-Channel Analyzer: 256 Channel MCA. 65535 counts/channel. RS-232 serial link can control MCA operation.
Temperature Dependence: less than ± 15 % from 15 to 45°C
Humidity Response: Less than 10 % change from 0 to 95 % RH non condensing.
Neutron Energy Response: 70 KeV to 20 MeV
Gamma Response: Less than 1% at 1 RAD/h.
Warm Up Time: 15 seconds
Battery Life: 100 hours; 6 ea alkaline C cells.
Serial RS232 link: In Rate and Integrate mode, sends radiation level on command. In MCA mode dumps MCA raw channel data, count time, total counts and
Instrument Calibration Factor. Baud rate is 9600. Available if the MCA option is
installed.