Download User Manual
Transcript
REM 500 INSTRUCTION MANUAL 11 DATA DUMP OF CHANNELS Sending a D followed by another character when the instrument is in the MCA mode will dump the data from the channels. The Data record is as follows: 256 lines of 5 digits. This is the counts for each channel. The counts are from 0 to 65534. 65535 is the maximum count even if the counts are more than this. There is no overflow indication other that keeping the count at the maximum value. The next is a line of counts and time. The counts are in HEX and are the total counts of all channels. This is 6 hex digits. The time is in HR:MIN:SEC. This line is the same as sent every second when the MCA is running. XV. PRINCIPAL OF OPERATION The Instrument is basically a 256 channel multichannel analyzer or MCA with a dedicated program coupled to a Tissue Equivalent Rossi type spherical proportional counter. 1. ELECTRONICS/SOFTWARE In normal operation the instrument gathers the pulses from the detector, measures their height, counts them and stores them according to their height, and then calculates the correct REM or RAD. Every 10 seconds it goes through this routine, and every 10 second interval is separate. In the rate mode it uses 1, 3, or 6 (de pending on the TC) of these 10 second intervals for the total value. In the integrate mode it just sums the results of each 10 second calculation. Thus every 10 second operation is the same regardless of the mode or the time constant. The display is derived from the formulas: For a 10 second integration period, where TC = 1 for 10 sec, 3 to 30 sec, 6 for 60 sec and 360 for integrate. REM = K 255 ∑× 5 RAD = K 255 ∑× 5 CHAN # × CNTS per chan × QF TC × 256 . CHAN # × CNTS per chan TC × 256 . The QF for the channels is derived from a smoothing of the ICRP values and ranges from 1 to 24.8. K is the Calibration Factor. If you wish a different QF then please consult HPI. The RAD display is thus the same as the REM except for the Quality factors. The average quality factor can be derived by dividing the REM reading by the RAD reading.