REM 500 INSTRUCTION MANUAL
11
DATA DUMP OF CHANNELS
Sending a D followed by another character when the instrument is in the MCA
mode will dump the data from the channels. The Data record is as follows:
256 lines of 5 digits. This is the counts for each channel. The counts are from 0 to
65534. 65535 is the maximum count even if the counts are more than this. There
is no overflow indication other that keeping the count at the maximum value. The
next is a line of counts and time. The counts are in HEX and are the total counts of
all channels. This is 6 hex digits. The time is in HR:MIN:SEC. This line is the
same as sent every second when the MCA is running.
XV. PRINCIPAL OF OPERATION
The Instrument is basically a 256 channel multichannel analyzer or MCA with a
dedicated program coupled to a Tissue Equivalent Rossi type spherical proportional counter.
1. ELECTRONICS/SOFTWARE
In normal operation the instrument gathers the pulses from the detector, measures
their height, counts them and stores them according to their height, and then calculates the correct REM or RAD. Every 10 seconds it goes through this routine,
and every 10 second interval is separate. In the rate mode it uses 1, 3, or 6 (de pending on the TC) of these 10 second intervals for the total value. In the integrate
mode it just sums the results of each 10 second calculation. Thus every 10 second
operation is the same regardless of the mode or the time constant.
The display is derived from the formulas:
For a 10 second integration period, where TC = 1 for 10 sec, 3 to 30 sec, 6 for 60
sec and 360 for integrate.
REM = K
255
∑×
5
RAD = K
255
∑×
5
CHAN # × CNTS per chan × QF
TC × 256
.
CHAN # × CNTS per chan
TC × 256
.
The QF for the channels is derived from a smoothing of the ICRP values and
ranges from 1 to 24.8. K is the Calibration Factor. If you wish a different QF then
please consult HPI. The RAD display is thus the same as the REM except for the
Quality factors. The average quality factor can be derived by dividing the REM
reading by the RAD reading.