1260 User Manual Publication No. 980673-999 Rev. A Table 3-1, 1260 Option 01 Status Byte Bit Number Description 1 (lsb) Unused 2 Unused 3 Unused 4 Scan List At Break Point 5 Ready 6 Programming Error 7 Service Requested 8 (msb) Unused The Programming error bit cannot be masked, and so the range of values for valid SRQMASKs is 32 - 63, 96 - 127, 160 - 191 and 223 - 255. The power up (default) value of the SRQMASK is 40, i.e., programming error or Scanlist at break point will cause SRQ to be asserted. STORE STORE <non-vol 1|2|.......|50 location> where <non-vol location> = This command will store the present state of all switching modules resident in the VXI chassis in the non-volatile memory location specified. A STORED configuration must be recalled into the same module set as was stored. Any changes in module location will be disregarded as long as the same module types are at the same module addresses as were stored. Operation 3-14 Astronics Test Systems