Download MetaFluor Analyst
Transcript
Advanced Image Processing background. The three parameter one (y=a*e^(-b*x)+c) serves more general purpose. Regression method, number of iterations, precision The Simplex algorithm works the most reliably. Regression method, number of iterations, precision should not be changed. Action • The difference of fitted function from its first value is subtracted from the frames • Frames are factorized by the relative change of fitted function from its first value Show Fit Calculates the regression, shows the parameters, and displays the results on the graph (in red). OK Calculates the regression and process the images sequence as specified at Action filed. Calibration Select Calibration from the Calculate menu to open the Calibration dialog box, as shown below: Grynkiewicz Equation to calibrate ratiometric (or non ratiometric) fluorescent dyes: C = Kd R − vRmin F f 2 vRmax − R Fb 2 Where the concetration (C) is calculated from the dye dissociation constant (Kd), the maximal and minimal ratio values (Rmax and Rmin), the intracellular viscosity (use 1 for water, usually between 0.6-0.8), and from the free and bound dye fluorescence at the wavelength in the denominator (Ff2 and Fb2). For non ratiometric calibration use Fmax and Fmin fluorescence intensities instead of Rmax and Rmin, taking Ff2 and Fb2 equal to 1. Normalization Proportional scaling for general purposes: C=k F − Fmin Fmax − Fmin Where the concetration (C) is calculated from a arbitrary constant (k), and the maximal and minimal fluorescent intesities (Fmax and Fmin). MetaFluor Analyst 6-3 Version 1.0.93