OB Measurements and Calculations Fetal Trunk Area (FTA) (continued) Two distances 1. Select FTA; an active caliper displays. 2. Make the first distance measurement: a. To position the active caliper at the start point, move the Trackball. b. To fix the start point, press Set. The system fixes the first caliper and displays a second active caliper. c. To position the second active caliper at the end point, move the Trackball. A dotted line connects the measurement points, if preset accordingly. d. To complete the measurement, press Set. The system displays the distance value in the Results Window and displays an active caliper. 3. To make the second distance measurement, repeat steps a–d above. The system displays the measurement in the Results Window. HINTS • Before you complete a measurement: • To toggle between active calipers, press Measure. • To erase the second caliper and the current data measured and start the measurement again, press Clear once. • To rotate through and activate previously fixed calipers, adjust Cursor Select. • After you complete the measurement, to erase all data that has been measured to this point, but not data entered onto worksheets, press Clear. LOGIQ Book Basic User Manual Direction 2321062-100 Rev. 5 9-27