The motorized stage will not function properly if these settings
are incorrect.
*Basically use same setting for (1) and (7)
The positions of the stage are calculated from (9) Pulses for one
rotation, (6) F-value multiplied by (7) Pulse Multiple Factor
sent out to the built-in driver and divided by (1) Micro-step
setting.
*Memorize the position after setting accomplish.
By setting the (7) Pulse Multipllication Factor and (1) Micro-step
setting the same, (9) Pulses for one rotation, (6) F-value and
(11) Backlash compensation are all multiplied by (7) Pulse
Multiple Factor and registered in the controller as position
parameter. If the value different from (1) Micro-step setting is
used, pulses per one rotation mismatches and cause incorrect
positioning.
Controller
・Position information
・Operation knob
(6) F-value
Built-in
(9) Pulse per rotation
(10) S-value
/ (1) Micro-step
x (7) Multiplier
division
(11) Backlash
Compensation
Fig. 3-2-4 Setting Diagram