Mechanics The Slave B PCB controls both the rinsing module and the stirrer module. When turning on power, perform the initial adjustment of the stirrer module first if this has not been done. 1. Adjusting cuvette vertical drive 1. Move the vertical shaft up and down before turning on the power. This also can be done after power is on if you are in the off-line mode and magnetization is turned off. Confirm that it moves smoothly. To disable magnetization (motor power) use the dip switch SW3 on the Slave B CPU. Set Dip Switch SW3 Bit 7 to ON. Depress the Reset switch. Set SW3 bit 3 to ON. Change position of SW3 bit 8 (start operation) Set SW3 to default positions when procedure is completed. 2. Adjust the module so the top face of the light shielding plate becomes flush with the top face of the upper photo sensor when the shaft is moved up until it reaches the upper stop position. Upper mechanical stop is set to 23.4 mm. Adjust the module also so the bottom of the light shielding plate becomes flush with the bottom of the lower photo sensor when the shaft is moved down and reaches the lower stop position. Lower mechanical stop is set to 23.3 mm. 2. Adjusting the rinse probe 1. Move the rinsing probe up and down by hand, make sure that they move smoothly. 2. Make sure that the rinsing probes ( 6 total ) spring back to their upper limits when pulled down by hand and released. 3. Make sure they are all straight and enter the reaction cuvettes without striking the sides, this excludes the drying tip which should slide smoothly down the cuvette wall. 3. Turning on the power and checking the sensors 1. This procedure places the instrument in the off line mode to provide sensor checking. 2. Turn on the power. Locate the Dip Switch SW3 on the Slave B PCB. Set Bit 7 to the ON position. Depress the reset switch. Set Bits 1 to 6 to the ON positions. Set Bit 8 to the ON position. 3. The sensor status is displayed by the LED indication. Check the LED indication table (4) Check Sensor Stirrer / Rinse. Insure all sensors are operation correctly. 4. Reset dip switch SW3 to default position. 4.38 Instrumentation Laboratory