4.8 VOC-TRAQ Maintenance
The VOC-TRAQ’s rugged, durable design provides for trouble-free operation over
the course of its lifetime. However, like all photo-ionization sensors, maintenance
may be required.
In a polluted environment, window contamination can degrade the sensor’s
performance. One indication of this problem is a loss of sensitivity. Another
possible effect is more noise in a VOC-TRAQ that was properly calibrated. The
sensor is still useful with a lower sensitivity but for best results clean the lamp
window. For lamp cleaning instructions refer to appendix III
4.9 VOC-TRAQ Disassembly
The VOC-TRAQ contains a piD-TECH® plus photoionization detector with a
10.6eV Lamp. Periodically the lamp in this sensor needs cleaning. The remove
the sensor unscrew the two caps from the VOC-TRAQ probe and remove the PCB
assemblies. The PID unplugs from the large rectangular printed circuit board
(PCB). Refer to appendix III for instructions on cleaning the lamp window.
Reassemble the VOC-TRAQ by installing the rectangular PCB into the base cap.
Reinsert the PID into the connector. Screw the tubular mid section on to the
base cap and the screw on the top cap. See photo below.