5) In the Status Window check the box to “Read ADS1100 Continuously”.
6) In the Main Window set “Supply Vcc (V)” to 5.0V
7) The “Rtest” potentiometer and “J4” on the PGA309 Sensor Interface Board will be used for this test. On the
PGA309 PC Interface Board the ADS1100 ADC will be used for this test. Table 4 below details the
actions and acceptable results for a pass condition of the PGA309 Sensor Interface and PGA309 PC
Interface Boards.
Step
Rtest
1
Rotate Fully Clockwise
(until audible clicking is heard)
Rotate Fully
Counter-Clockwise
(until audible clicking is heard)
Don’t Care
Replace J4
2
3
4
J4
Position
Vexc
ADS1100 read
(Volts min)
0.19
ADS1100 read
(Volts max)
0.26
Vexc
3.9
4.1
Open
Vexc
4.9
Don’t Care
5.0
Don’t Care
Table 4: Hardware Checkout Voltages and Settings
8) In the Main Window, under “PGA309 Temp ADC”, click on “Convert” (see Figure 16).
9) Temperature reading in °C should be the PGA309 ambient +/-2°C. Readout will be in Main Window under
“PGA309 Temp ADC” section as shown in Figure 16.
Figure 16: Internal Temperature Reading
If all steps above, in Section 6, have been successfully completed then the PGA309 Sensor Interface Board,
PGA309 PC Interface Board, PC Cable, PGA309DK Board Control Software are all operating correctly.