Download PGA309EVM User`s Guide (Rev. D
Transcript
4) The remainder of the worksheets contain the templates of the curve-fitting algorithms and back-ground calculation data 9.4 Step-by-Step Calibration Using PGA309 Designer’s Kit. 1) The PGA309EVM Sensor Interface Board with a sensor or a custom module are initially configured as described in Section 8.2. It is measured at the first (usually the room) temperature, Tmeas1, and at either 2 or 3 pressure levels (see Section 8.1). The output should be within the adjustment ranges of the Zero and Gain DACs. It is important to enable the EXTEN and INTEN bits and/or read back the Alarm Status (Register 8) to see if any of the external or internal comparators raised a flag indicating either sensor fault or an internal signal path saturation. In case of a fault or an unsatisfactory output signal range the coarse settings must be adjusted and the measurement repeated. After a valid output range has been achieved, the following settings must remain fixed: § Reference and Excitation Level § Coarse RTI Offset, § Coarse Front-End and Output Amplifier Gains, § Temp ADC Configuration, etc. 2) The current configuration must be entered into the PGA309 Calibration spreadsheet. To do this, perform the following actions: A) Open the PGA309 Calibration Spreadsheet (click “Enable Macros” if asked). Upon opening you will be asked if you would like to open the “PGA309EVM Board Control Program”. Answer “Yes”. If you already have the software running, a new window will not be open, but a data link will be established. B) Make sure the current PGA309 register values are displayed in the Board Control Program (click “Read PGA309” button – see Figure 38). Figure 38: Board Control Software Main Window “Read PGA” Button C) Return to the “Main” Worksheet and click “Load Registers from Bench”. The register values in the cells B3:B12 will be updated. The values of the gain, offset and other parameters are set on the next worksheet accordingly to the PGA309 register configuration. D) Click “Prepare Calibration Sheet”. The dialog box in Figure 39 will appear. Select the appropriate algorithm. The script will copy the right template to the “Sensor Curvefit” Worksheet and open the next page. E) Select the appropriate Temperature Calibration Method (see Figure 40). Page 44 of 62