Download Tip Adapters - New Objective, Inc.
Transcript
PicoTip 7. 8. 9. 10. ® Adapter Instruction Manual end of the fused-silica should protrude slightly (approximately 2 mm) beyond the sleeve ensuring even positioning of the silica and sleeve within the union. Finger-tighten the white gauge plug into one side of the union or 1/16” adapter (see Figure 23). Screw the assembled MicroTight™ fitting with clamping mechanism into the 6-32 end of the adapter, making sure that the tubing and sleeve are flush against the gauge plug. Do not over tighten (Figure 23). Remove the gauge plug. Into the other end of the fitting, insert the transfer line with appropriate sleeve and fitting Figure 24). The sleeve and fitting had been previously threaded through the ESI manifold on the ADPT-TLC/ADPT-TLD. Push the tubing flush with the distal end of the PicoTip® and tighten both fittings. You should now have a zero-dead volume connection. Over tightening can both damage the silica tubing and prevent the fittings from being reused. When properly fitted removal of the fitting would reveal the PicoTip to be flush with the PEEK™ sleeve. Both the PicoTip and sleeve should protrude from the end of the fitting (Figure 25). 6.2 Mounting on the Instrument 1. You are now ready to mount the tip assembly on the instrument. On the LCQ™ & TSQ®, push the assembly as far into the mounting adapter as possible and tighten set screw. If the tip protrudes too far from the mount, it may contact the capillary inlet. This situation will damage the tip and possibly the inlet. On the Protana, simply slide the assembly into the mounting hole and tighten the set screw. 2. Connect the transfer line to a source of liquid and pressurize the system as required (syringe pump at 100 nl/min etc.). After the line is primed, you should see a droplet of liquid at the end of the PicoTip. Inspect the union for any leaks etc., and tighten if necessary. 3. For LCQ/TSQ only – Remove the vent cap and slide the ESI probe to its forward limit. Pull it back and inspect the tip. If it is not broken or chipped, then you are a safe distance from the capillary inlet. Provided that all of the components are 14 Figure 23 Figure 24 Distal end of tip should be flush with the sleeve Figure 25