7.
8.
9.
10.
®
Adapter
Instruction
Manual
end of the fused-silica should protrude slightly
(approximately 2 mm) beyond the sleeve ensuring
even positioning of the silica and sleeve within
the union.
Finger-tighten the white gauge plug into one side
of the union or 1/16” adapter (see Figure 23).
Screw the assembled MicroTight™ fitting with
clamping mechanism into the 6-32 end of the
adapter, making sure that the tubing and sleeve
are flush against the gauge plug. Do not over
tighten (Figure 23).
Remove the gauge plug. Into the other end of the
fitting, insert the transfer line with appropriate
sleeve and fitting Figure 24). The sleeve and fitting
had been previously threaded through the ESI
manifold on the ADPT-TLC/ADPT-TLD. Push the
tubing flush with the distal end of the PicoTip®
and tighten both fittings. You should now have a
zero-dead volume connection. Over tightening
can both damage the silica tubing and prevent the
fittings from being reused.
When properly fitted removal of the fitting would
reveal the PicoTip to be flush with the PEEK™
sleeve. Both the PicoTip and sleeve should
protrude from the end of the fitting (Figure 25).
6.2 Mounting on the Instrument
1. You are now ready to mount the tip assembly
on the instrument. On the LCQ™ & TSQ®, push
the assembly as far into the mounting adapter
as possible and tighten set screw. If the tip
protrudes too far from the mount, it may contact
the capillary inlet. This situation will damage the
tip and possibly the inlet. On the Protana, simply
slide the assembly into the mounting hole and
tighten the set screw.
2. Connect the transfer line to a source of liquid and
pressurize the system as required (syringe pump
at 100 nl/min etc.). After the line is primed, you
should see a droplet of liquid at the end of the
PicoTip. Inspect the union for any leaks etc., and
tighten if necessary.
3. For LCQ/TSQ only – Remove the vent cap and
slide the ESI probe to its forward limit. Pull it back
and inspect the tip. If it is not broken or chipped,
then you are a safe distance from the capillary
inlet. Provided that all of the components are
14
Figure 23
Figure 24
Distal end of tip
should be flush
with the sleeve
Figure 25