Download EASY-Spray Series Ion Source User Guide
Transcript
4 Connecting the LC to the EASY-Spray Column Connecting the LC to the EASY-Spray Column To plumb a two-column setup from an RSLCnano instrument 1. Connect the precolumn to ports 2 and 5 on one of the switching valves (Figure 23). 2. Insert the nanoViper fitting on the 20 μm ID nanoViper capillary into port 4 on the same switching valve. 3. Insert the fitting on the other end of the capillary into the Viper union, tighten the fitting by hand until you feel resistance, and then tighten the fitting again by an additional 1/8- to 1/4-turn (45–90 degrees). Do not tighten the fitting by more than a 1/4-turn. Note See the caution statement on page 28. 4. Connect the union to the EASY-Spray column to complete the two-column setup (Figure 23). Figure 23. Example of a two-column setup for the RSLCnano system (preconcentration) EASY-Spray column Viper union 20 μm ID 20 μm ID Column compartment with two switching valves Precolumn 30 EASY-Spray Ion Source User Guide Thermo Scientific