Download Model 48i - User Manual
Transcript
C-Link Protocol Commands Datalogging diag volt mib This command reports the diagnostic voltage measurements on the measurement interface board. The sequence of voltages is: Positive 24, positive 15, negative 15, positive 5, positive 3.3, positive 18 IR, positive 18 MOT, and VBIAS. Each voltage value is separated by a space. Send: Receive: diag volt mib diag volt mib 24.1 14.9 -14.9 4.9 3.2 17.9 17.9 diag volt iob This command reports the diagnostic voltage measurements on the I/O expansion board. The sequence of voltages is: Positive 24, positive 5, positive 3.3, and negative 3.3. Each voltage value is separated by a space. Send: Receive: diag volt iob diag volt iob 24.1 4.9 3.2 -3.2 version This command reports the version of all the firmware components. The example that follows shows a list of firmware components that were displayed by issuing the version command. The components displayed will vary depending on the configuration of the instrument. Send: Receive: Datalogging version version Program = 01.05.79.225 Library = 01.01.60.167 Kernel = 2.4.24-uc0-003-Thermo Board = 81, File = /usr/application.hex Board App = 11.3.100 Bl 4.0.97 File App = 11.3.100 Bl 4.0.97 Board = 84, File = /usr/application.hex Board App = 11.3.100 Bl 4.0.97 File App = 11.3.100 Bl 4.0.97 Arc Bl = 170711 clr records This command will clear all lrecs and srecs that have been saved. Send: Receive: clr records clr records ok set clr lrecs set clr srecs These commands will clear only the lrecs or only the srecs that have been saved. The following example clears srecs. Thermo Fisher Scientific Model 48i Instruction Manual B-19