Download 6200E - Teledyne Analytical Instruments
Transcript
Model 6200E Instruction Manual Operating Instructions 6.10.10. COM Port Testing The serial ports can be tested for correct connection and output in the COMM menu. This test sends a string of 256 ‘w’ characters to the selected COM port. While the test is running, the red LED on the rear panel of the analyzer should flicker. To initiate the test press the following key sequence. SAMPLE RANGE = 500.0 PPB H2S =XXX.X < TST TST > CAL SAMPLE 8 SETUP X.X SETUP ENTER SETUP PASS : 818 1 PRIMARY SETUP MENU CFG DAS RNGE PASS CLK MORE SETUP X.X ID COM1 COMMUNICATIONS MENU COM2 SETUP X.X ENTR EXIT 8 SETUP X.X SET> SETUP X.X EXIT <SET SET> EXIT Select which COM port to test. COM1 MODE:0 EDIT EXIT COM1 BAUD RATE:19200 EDIT EXIT SECONDARY SETUP MENU SETUP X.X COMM VARS DIAG COM1 : TEST PORT EXIT <SET Test runs automatically SETUP X.X <SET TEST EXIT TRANSMITTING TO COM1 TEST EXIT returns to COMM menu EXIT 6.11. Using the Data Acquisition System (iDAS) The M6200E analyzer contains a flexible and powerful, internal data acquisition system (iDAS) that enables the analyzer to store concentration and calibration data as well as a host of diagnostic parameters. The iDAS of the M6200E can store up to about one million data points, which can, depending on individual configurations, cover days, weeks or months of valuable measurements. The data are stored in non-volatile memory and are retained even when the instrument is powered off. Data are stored in plain text format for easy retrieval and use in common data analysis programs (such as spreadsheet-type programs). The iDAS is designed to be flexible, users have full control over the type, length and reporting time of the data. The iDAS permits users to access stored data through the instrument’s front panel or its communication ports. Using APICOM, data can even be retrieved automatically to a remote computer for further processing. The principal use of the iDAS is logging data for trend analysis and predictive diagnostics, which can assist in identifying possible problems before they affect the functionality of the analyzer. The secondary use is for data analysis, documentation and archival in electronic format. To support the iDAS functionality, Teledyne Analytical Instruments offers APICOM, a program that provides a visual interface for remote or local setup, configuration and data retrieval of the iDAS M6200E Rev: A1 119