Download ECG Scanning Software

Transcript
4 Lifescreen Operating Instructions
Good Recording Quality
Good recording quality is your first priority
4.1.1
Hook-up
Proper hook-up, or patient connection, is essential to ensure intelligible ECG
recordings. Inadequate skin preparation will result in excessive artefact in the
signal, causing measurement of false intervals and difficulty in visual
interpretation.
4.1.2
Skin Preparation
There are 3 main steps in good skin preparation:
CLEAN the skin using isopropyl alcohol, soapy water or a non-alcoholic swab.
This removes grease, which prevents the electrode adhering and impairs
conductivity.
SHAVE the site as needed with a new disposable razor - the electrode will not
remain in place on hair!
RUB the site with dry gauze to lightly redden the skin. This removes dry, dead
skin, lowering impedance and reducing pick-up of interference.
NOTE
Take care not to break the skin or cause bleeding. Avoid
abrasion (rubbing) on patients with delicate skin,
including children and elderly patients.
4.1.3
ECG Electrodes
Always use good quality electrodes designed for long-term (Holter) recording, and
ensure adequate skin preparation. We recommend Ambu Blue Sensor VL-00-S
for 24 hour recordings, and VLC-00-S for extended recordings up to 7 days.
Further information can be found at www.ambu.com.
Apply the electrodes in the positions shown in the ECG recorder instruction
manual.
Check the ECG quality on the recorder, as shown in the recorder instruction
manual.
Lifescreen Instruction Manual