Download Agilent Technologies E3632A Technical data
Transcript
Service Electrostatic Discharge (ESD) Precautions 2 Electrostatic Discharge (ESD) Precautions Almost all electrical components can be damaged by electrostatic discharge (ESD) during handling. Component damage can occur at electrostatic discharge voltages as low as 50 V. The following guidelines will help prevent ESD damage when serving the power supply or any electronic device. • Disassemble instruments only in a static- free work area. • Use a conductive work area to dissipate static charge. • Use a conductive wrist strap to dissipate static charge accumulation. • Minimize handling. • Keep replacement parts in original static- free packaging. • Remove all plastic, styrofoam, vinyl, paper, and other static- generating materials from the immediate work area. • Use only anti- static solder suckers. Surface Mount Repair Surface mount components should only be removed using soldering irons or desoldering stations expressly designed for surface mount components. NOTE E3632A Service Guide Use of conventional solder removal equipment will almost always result in permanent damage to the printed circuit board and will void your Agilent factory warranty. 51